PENILE IMPLANT COMPONENTS
Penile Cylinders
The cylinders of the penile implant consist of three layers. The inner and outer layers are made of Silicone. The middle layer, orfabric layer, is made from a Dacron and Lycra weave.
The middle Dacron layer is designed to prevent uneven expansion of the cylinders. Before this layer was introduced, cylinder aneurysm or bubble formation was a problem. The Dacron weave effectively eliminated this problem of uneven cylinder expansion.
Reservoir
The reservoir is composed of silicone and is coated on the inside surface with Paralyne. It is a seamless container and leakage is a rare event. The reservoir comes in a spherical shape for placement next to the urinary bladder. The reservoir also comes in a flat shape. This is called the “Conceal Reservoir” and it is designed for placement behind the rectus muscle in the wall of the abdomen.
Pump
The compression pump has two functions. The first function is to return the penis to the flaccid state. The release valve on top of the pump allows water to drain from the cylinders and this results in a flaccid penis. This button needs to be compressed for only two seconds. Consequently, this button is termed the “Momentary Squeeze”.
The second function of the pump is to produce a rigid penis. Compression of the pump bulb transfers water to the cylinders. This produces a rigid penis. The pump is designed to transfer water with a minimum number of compressions.
Finally, the pump contains a lockout valve. This valve prevents auto-inflation of the cylinders. Auto-inflation results in an erection due to the spontaneous flow of water from the reservoir to the cylinders. This can produce an embarrassing situation.
Tubing
The tubing that connects the components has been reinforced. It is designed to be kink-resistant.If the tubing becomes kinked, it will prevent the implant from either inflating or deflating. Also, the new reinforced tubing rarely leaks.
PENILE IMPLANT TYPES
The AMS 700 penile implant comes in different models based upon how the cylinders are constructed. The middle layer of the cylinders consist of a fabric layer made from Dacron and Lycra. The Dacron layer can be weaved in a uni-directional or bi-directional fashion. The uni-directional weave allows the cylinder to expand in girth during the erection. The bi-directional weave allows expansion of the cylinder in length and girth.
AMS 700 CX
CX stands for controlled expansion. Upon inflation, the cylinders expands in girth. In other words, the cylinders becomes thicker during the erection. This is a result of the uni – directional Dacron weave in the middle layer of the cylinder.
AMS 700 LGX
LGX stands for length and girth expansion. Upon inflation, the cylinders expand in girth and in length. This is the result of the bi – directional Dacron weave in the middle layer of the cylinders.
AMS 700 CXR
CXR stands for controlled expansion restricted. This cylinder is narrow and is designed for the short or narrow penis. It is also ideal for patients with scarring of the penis due to Peyronie’s disease or scarring due to prior implant infection.
IMPLANT SIZES
The implants come in different lengths including 12 cm, 15 cm, 18 cm, and 21 cm. The penis is carefully sized during surgery and the largest implant that the penis can accommodate is selected.
Rear Tip Extenders
Rear tip extenders are conical extenders that attach to the rear of the implant. They range in size from 1 to 6 cm. They effectively make the implant longer and allow for more precise sizing of the implant.
INHIBIZONE
In order to decrease the rate of penile implant infection, the AMS 700 is coated with a material called Inhibizone. Inhibizone consists of two antibiotics – Rifampin and Minocycline. These two antibiotics are attached to the surface of the implant and are released very slowly following surgery. It is important to know ahead of surgery if the patient has an allergy to either Rifampin, Minocycline, or to Tetracycline. Inhibizone has cut the rate of penile implant infection in half.
HOW DOES THE PENILE IMPLANT WORK?
The patient compresses the pump in the scrotum and water is transferred to the cylinders. This confers rigidity to the penis. Compression of the release valve, which is located just above the pump, allows water to drain from the cylinders. This results in a flaccid penis.
PENILE IMPLANT COMPONENTS
Penile Cylinders
The cylinders of the penile implant consist of three layers. The inner and outer layers are made of Silicone. The middle layer, orfabric layer, is made from a Dacron and Lycra weave.
The middle Dacron layer is designed to prevent the uneven expansion of the cylinders. Before this layer was introduced, cylinder aneurysm or bubble formation was a problem. The Dacron weave effectively eliminated this problem of uneven cylinder expansion.
Despite these improvements in design, cylinder leakage and cylinder malfunction remained a problem. To address this issue, AMS introduced a new design modification. Starting in 2001, the AMS implant was coated with a polymer called Paralyne. This substance reduced the friction between the Silicone layers and the Dacron weave. Paralyne has successfully reduced the rate of cylinder leakage and cylinder malfunction.
All of these changes have improved the longevity of the penile implant. Now, the penile implant has a functional lifetime of 15 years on the average. The penile implant is currently one of the most reliable and long lasting devices that is implanted into humans.
Reservoir
The reservoir is composed of silicone and is coated on the inside surface with Paralyne. It is a seamless container and leakage is a rare event. The reservoir comes in a spherical shape for placement next to the urinary bladder. The reservoir also comes in a flat shape. This is called the “Conceal Reservoir” and it is designed for placement behind the rectus muscle in the wall of the abdomen.
Pump
The compression pump has two functions. The first function is to return the penis to the flaccid state. The release valve on top of the pump allows water to drain from the cylinders and this results in a flaccid penis. This button needs to be compressed for only two seconds. Consequently, this button is termed the “Momentary Squeeze”.
The second function of the pump is to produce a rigid penis. Compression of the pump bulb transfers water to the cylinders. This produces a rigid penis. The pump is designed to transfer water with a minimum number of compressions.
Finally, the pump contains a lockout valve. This valve prevents auto-inflation of the cylinders. Auto-inflation results in an erection due to the spontaneous flow of water from the reservoir to the cylinders. This can produce an embarrassing situation.
Tubing
The tubing that connects the components has been reinforced. It is designed to be kink resistant. If the tubing becomes kinked, it will prevent the implant from either inflating or deflating. Also, the new reinforced tubing rarely leaks.
INHIBIZONE
In order to decrease the rate of penile implant infection, the AMS 700 is coated with a material called Inhibizone. Inhibizone consists of two antibiotics – Rifampin and Minocycline. These two antibiotics are attached to the surface of the implant and they are released very slowly following surgery. It is important to know ahead of surgery if the patient has an allergy to either Rifampin, Minocycline or to Tetracycline. Inhibizone has cut the rate of penile implant infection in half.
What procedures are available at Group Florence Nightingale Hospitals?
How many medical staff are there at Group Florence Nightingale Hospitals and what accreditations do they have?
Penile prosthesis or penile implant surgery is performed as a treatment for men with erectile dysfunction and in some cases to increase the size of the penis (Penis Augmentation). These devices are either malleable (bendable) or inflatable. The simplest type of prosthesis consists of a pair of malleable rods surgically implanted within the erection chambers of the penis, under general anesthetic. With this type of implant, the penis is always semi-rigid and merely needs to be lifted or adjusted into the erect position to initiate sex. This type of implant is a good choice for men with spinal cord injuries and/or limited hand strength. Today, many men choose a hydraulic, inflatable prosthesis, which allows them to have an erection when they choose, and it's easier to conceal. It is also more natural.
A penile implant is usually used when there is a clear medical cause for erectile dysfunction and when the problem is unlikely to resolve or improve naturally or with other medical treatments. Sometimes a penile prosthesis is implanted during surgery to reconstruct the penis when scarring has caused erections to curve (Peyronie's disease).
After the surgery, you will likely need to take medications to help ease the pain and you may feel mild pain for several weeks. Some swelling and bruising may occur for the first few days. You will be advised to avoid water in the penis areas for at least a week.
Most patients can go back to strenuous physical activity and sexual activity within four to six weeks. It is best to ask your surgeon when you can resume your normal activity and sexual intercourse. During your recovery period, you will have to take antibiotics to prevent infection and the incisions will heal within five days following surgery.
A penile prosthesis surgery has a high success rate, with about 80% to 90% of men satisfied with the results. Around 90% to 95% of inflatable prosthesis produce erections suitable for intercourse. 60% to 80% of implant devices will continue to work for ten years.
For an in-depth analysis of the pros and cons of Penile prosthesis, watch this short video.
PENILE IMPLANT COMPONENTS
Penile Cylinders
The cylinders of the penile implant consist of three layers. The inner and outer layers are made of Silicone. The middle layer, orfabric layer, is made from a Dacron and Lycra weave.
The middle Dacron layer is designed to prevent uneven expansion of the cylinders. Before this layer was introduced, cylinder aneurysm or bubble formation was a problem. The Dacron weave effectively eliminated this problem of uneven cylinder expansion.
Reservoir
The reservoir is composed of silicone and is coated on the inside surface with Paralyne. It is a seamless container and leakage is a rare event. The reservoir comes in a spherical shape for placement next to the urinary bladder. The reservoir also comes in a flat shape. This is called the “Conceal Reservoir” and it is designed for placement behind the rectus muscle in the wall of the abdomen.
Pump
The compression pump has two functions. The first function is to return the penis to the flaccid state. The release valve on top of the pump allows water to drain from the cylinders and this results in a flaccid penis. This button needs to be compressed for only two seconds. Consequently, this button is termed the “Momentary Squeeze”.
The second function of the pump is to produce a rigid penis. Compression of the pump bulb transfers water to the cylinders. This produces a rigid penis. The pump is designed to transfer water with a minimum number of compressions.
Finally, the pump contains a lockout valve. This valve prevents auto-inflation of the cylinders. Auto-inflation results in an erection due to the spontaneous flow of water from the reservoir to the cylinders. This can produce an embarrassing situation.
Tubing
The tubing that connects the components has been reinforced. It is designed to be kink-resistant.If the tubing becomes kinked, it will prevent the implant from either inflating or deflating. Also, the new reinforced tubing rarely leaks.
PENILE IMPLANT TYPES
The AMS 700 penile implant comes in different models based upon how the cylinders are constructed. The middle layer of the cylinders consist of a fabric layer made from Dacron and Lycra. The Dacron layer can be weaved in a uni-directional or bi-directional fashion. The uni-directional weave allows the cylinder to expand in girth during the erection. The bi-directional weave allows expansion of the cylinder in length and girth.
AMS 700 CX
CX stands for controlled expansion. Upon inflation, the cylinders expands in girth. In other words, the cylinders becomes thicker during the erection. This is a result of the uni – directional Dacron weave in the middle layer of the cylinder.
AMS 700 LGX
LGX stands for length and girth expansion. Upon inflation, the cylinders expand in girth and in length. This is the result of the bi – directional Dacron weave in the middle layer of the cylinders.
AMS 700 CXR
CXR stands for controlled expansion restricted. This cylinder is narrow and is designed for the short or narrow penis. It is also ideal for patients with scarring of the penis due to Peyronie’s disease or scarring due to prior implant infection.
IMPLANT SIZES
The implants come in different lengths including 12 cm, 15 cm, 18 cm, and 21 cm. The penis is carefully sized during surgery and the largest implant that the penis can accommodate is selected.
Rear Tip Extenders
Rear tip extenders are conical extenders that attach to the rear of the implant. They range in size from 1 to 6 cm. They effectively make the implant longer and allow for more precise sizing of the implant.
INHIBIZONE
In order to decrease the rate of penile implant infection, the AMS 700 is coated with a material called Inhibizone. Inhibizone consists of two antibiotics – Rifampin and Minocycline. These two antibiotics are attached to the surface of the implant and are released very slowly following surgery. It is important to know ahead of surgery if the patient has an allergy to either Rifampin, Minocycline, or to Tetracycline. Inhibizone has cut the rate of penile implant infection in half.
HOW DOES THE PENILE IMPLANT WORK?
The patient compresses the pump in the scrotum and water is transferred to the cylinders. This confers rigidity to the penis. Compression of the release valve, which is located just above the pump, allows water to drain from the cylinders. This results in a flaccid penis.
PENILE IMPLANT COMPONENTS
Penile Cylinders
The cylinders of the penile implant consist of three layers. The inner and outer layers are made of Silicone. The middle layer, orfabric layer, is made from a Dacron and Lycra weave.
The middle Dacron layer is designed to prevent the uneven expansion of the cylinders. Before this layer was introduced, cylinder aneurysm or bubble formation was a problem. The Dacron weave effectively eliminated this problem of uneven cylinder expansion.
Despite these improvements in design, cylinder leakage and cylinder malfunction remained a problem. To address this issue, AMS introduced a new design modification. Starting in 2001, the AMS implant was coated with a polymer called Paralyne. This substance reduced the friction between the Silicone layers and the Dacron weave. Paralyne has successfully reduced the rate of cylinder leakage and cylinder malfunction.
All of these changes have improved the longevity of the penile implant. Now, the penile implant has a functional lifetime of 15 years on the average. The penile implant is currently one of the most reliable and long lasting devices that is implanted into humans.
Reservoir
The reservoir is composed of silicone and is coated on the inside surface with Paralyne. It is a seamless container and leakage is a rare event. The reservoir comes in a spherical shape for placement next to the urinary bladder. The reservoir also comes in a flat shape. This is called the “Conceal Reservoir” and it is designed for placement behind the rectus muscle in the wall of the abdomen.
Pump
The compression pump has two functions. The first function is to return the penis to the flaccid state. The release valve on top of the pump allows water to drain from the cylinders and this results in a flaccid penis. This button needs to be compressed for only two seconds. Consequently, this button is termed the “Momentary Squeeze”.
The second function of the pump is to produce a rigid penis. Compression of the pump bulb transfers water to the cylinders. This produces a rigid penis. The pump is designed to transfer water with a minimum number of compressions.
Finally, the pump contains a lockout valve. This valve prevents auto-inflation of the cylinders. Auto-inflation results in an erection due to the spontaneous flow of water from the reservoir to the cylinders. This can produce an embarrassing situation.
Tubing
The tubing that connects the components has been reinforced. It is designed to be kink resistant. If the tubing becomes kinked, it will prevent the implant from either inflating or deflating. Also, the new reinforced tubing rarely leaks.
INHIBIZONE
In order to decrease the rate of penile implant infection, the AMS 700 is coated with a material called Inhibizone. Inhibizone consists of two antibiotics – Rifampin and Minocycline. These two antibiotics are attached to the surface of the implant and they are released very slowly following surgery. It is important to know ahead of surgery if the patient has an allergy to either Rifampin, Minocycline or to Tetracycline. Inhibizone has cut the rate of penile implant infection in half.
What procedures are available at Group Florence Nightingale Hospitals?
How many medical staff are there at Group Florence Nightingale Hospitals and what accreditations do they have?
PENILE IMPLANT COMPONENTS
Penile Cylinders
The cylinders of the penile implant consist of three layers. The inner and outer layers are made of Silicone. The middle layer, orfabric layer, is made from a Dacron and Lycra weave.
The middle Dacron layer is designed to prevent uneven expansion of the cylinders. Before this layer was introduced, cylinder aneurysm or bubble formation was a problem. The Dacron weave effectively eliminated this problem of uneven cylinder expansion.
Reservoir
The reservoir is composed of silicone and is coated on the inside surface with Paralyne. It is a seamless container and leakage is a rare event. The reservoir comes in a spherical shape for placement next to the urinary bladder. The reservoir also comes in a flat shape. This is called the “Conceal Reservoir” and it is designed for placement behind the rectus muscle in the wall of the abdomen.
Pump
The compression pump has two functions. The first function is to return the penis to the flaccid state. The release valve on top of the pump allows water to drain from the cylinders and this results in a flaccid penis. This button needs to be compressed for only two seconds. Consequently, this button is termed the “Momentary Squeeze”.
The second function of the pump is to produce a rigid penis. Compression of the pump bulb transfers water to the cylinders. This produces a rigid penis. The pump is designed to transfer water with a minimum number of compressions.
Finally, the pump contains a lockout valve. This valve prevents auto-inflation of the cylinders. Auto-inflation results in an erection due to the spontaneous flow of water from the reservoir to the cylinders. This can produce an embarrassing situation.
Tubing
The tubing that connects the components has been reinforced. It is designed to be kink-resistant.If the tubing becomes kinked, it will prevent the implant from either inflating or deflating. Also, the new reinforced tubing rarely leaks.
PENILE IMPLANT TYPES
The AMS 700 penile implant comes in different models based upon how the cylinders are constructed. The middle layer of the cylinders consist of a fabric layer made from Dacron and Lycra. The Dacron layer can be weaved in a uni-directional or bi-directional fashion. The uni-directional weave allows the cylinder to expand in girth during the erection. The bi-directional weave allows expansion of the cylinder in length and girth.
AMS 700 CX
CX stands for controlled expansion. Upon inflation, the cylinders expands in girth. In other words, the cylinders becomes thicker during the erection. This is a result of the uni – directional Dacron weave in the middle layer of the cylinder.
AMS 700 LGX
LGX stands for length and girth expansion. Upon inflation, the cylinders expand in girth and in length. This is the result of the bi – directional Dacron weave in the middle layer of the cylinders.
AMS 700 CXR
CXR stands for controlled expansion restricted. This cylinder is narrow and is designed for the short or narrow penis. It is also ideal for patients with scarring of the penis due to Peyronie’s disease or scarring due to prior implant infection.
IMPLANT SIZES
The implants come in different lengths including 12 cm, 15 cm, 18 cm, and 21 cm. The penis is carefully sized during surgery and the largest implant that the penis can accommodate is selected.
Rear Tip Extenders
Rear tip extenders are conical extenders that attach to the rear of the implant. They range in size from 1 to 6 cm. They effectively make the implant longer and allow for more precise sizing of the implant.
INHIBIZONE
In order to decrease the rate of penile implant infection, the AMS 700 is coated with a material called Inhibizone. Inhibizone consists of two antibiotics – Rifampin and Minocycline. These two antibiotics are attached to the surface of the implant and are released very slowly following surgery. It is important to know ahead of surgery if the patient has an allergy to either Rifampin, Minocycline, or to Tetracycline. Inhibizone has cut the rate of penile implant infection in half.
HOW DOES THE PENILE IMPLANT WORK?
The patient compresses the pump in the scrotum and water is transferred to the cylinders. This confers rigidity to the penis. Compression of the release valve, which is located just above the pump, allows water to drain from the cylinders. This results in a flaccid penis.
PENILE IMPLANT COMPONENTS
Penile Cylinders
The cylinders of the penile implant consist of three layers. The inner and outer layers are made of Silicone. The middle layer, orfabric layer, is made from a Dacron and Lycra weave.
The middle Dacron layer is designed to prevent the uneven expansion of the cylinders. Before this layer was introduced, cylinder aneurysm or bubble formation was a problem. The Dacron weave effectively eliminated this problem of uneven cylinder expansion.
Despite these improvements in design, cylinder leakage and cylinder malfunction remained a problem. To address this issue, AMS introduced a new design modification. Starting in 2001, the AMS implant was coated with a polymer called Paralyne. This substance reduced the friction between the Silicone layers and the Dacron weave. Paralyne has successfully reduced the rate of cylinder leakage and cylinder malfunction.
All of these changes have improved the longevity of the penile implant. Now, the penile implant has a functional lifetime of 15 years on the average. The penile implant is currently one of the most reliable and long lasting devices that is implanted into humans.
Reservoir
The reservoir is composed of silicone and is coated on the inside surface with Paralyne. It is a seamless container and leakage is a rare event. The reservoir comes in a spherical shape for placement next to the urinary bladder. The reservoir also comes in a flat shape. This is called the “Conceal Reservoir” and it is designed for placement behind the rectus muscle in the wall of the abdomen.
Pump
The compression pump has two functions. The first function is to return the penis to the flaccid state. The release valve on top of the pump allows water to drain from the cylinders and this results in a flaccid penis. This button needs to be compressed for only two seconds. Consequently, this button is termed the “Momentary Squeeze”.
The second function of the pump is to produce a rigid penis. Compression of the pump bulb transfers water to the cylinders. This produces a rigid penis. The pump is designed to transfer water with a minimum number of compressions.
Finally, the pump contains a lockout valve. This valve prevents auto-inflation of the cylinders. Auto-inflation results in an erection due to the spontaneous flow of water from the reservoir to the cylinders. This can produce an embarrassing situation.
Tubing
The tubing that connects the components has been reinforced. It is designed to be kink resistant. If the tubing becomes kinked, it will prevent the implant from either inflating or deflating. Also, the new reinforced tubing rarely leaks.
INHIBIZONE
In order to decrease the rate of penile implant infection, the AMS 700 is coated with a material called Inhibizone. Inhibizone consists of two antibiotics – Rifampin and Minocycline. These two antibiotics are attached to the surface of the implant and they are released very slowly following surgery. It is important to know ahead of surgery if the patient has an allergy to either Rifampin, Minocycline or to Tetracycline. Inhibizone has cut the rate of penile implant infection in half.
What procedures are available at Group Florence Nightingale Hospitals?
How many medical staff are there at Group Florence Nightingale Hospitals and what accreditations do they have?
CONTACT SUCCESSFUL